Model Number 42-9551 |
Device Problem
Material Fragmentation (1261)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 12/14/2021 |
Event Type
malfunction
|
Manufacturer Narrative
|
The sample is reported to be available, but has not yet been received by the manufacturer.A review of the device history record is not possible as no lot number was provided.All information reasonably known as of 28 dec 2021 has been included in this health authority report.Should additional information be obtained, a follow-up health authority report will be provided.The information provided by avanos medical inc.Represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to avanos medical inc.Avanos medical inc.Has no independent knowledge of the event reported but is relaying the information that was provided by the user facility where the incident occurred.This product incident is documented in the avanos medical complaint database and identified as complaint (b)(4).
|
|
Event Description
|
It was reported that when the nasal-jejunal tube was removed, the tip was noted to be missing.The device had been placed on (b)(6) 2021 under video fluroscopy.Per additional information received 17 dec 2021, the patient's condition is ok and no medical intervention was required.Per additional information received 20 dec 2021, the reporter confirmed that the tip was not removed from the patient.
|
|
Manufacturer Narrative
|
Correction: based on sample received and communications with the original reporter, the d4 model number and udi have been corrected.All other product information fields remain unaffected.All information reasonably known as of 23 feb 2022 has been included in this health authority report.Should additional information be obtained, a follow-up health authority report will be provided.The information provided by avanos medical inc.Represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to avanos medical inc.Avanos medical inc.Has no independent knowledge of the event reported but is relaying the information that was provided by the user facility where the incident occurred.This product incident is documented in the avanos medical complaint database and identified as complaint (b)(4).This information is submitted pursuant to 21cfr803, in compliance with the medical device reporting requirement and should not be considered to be an admission that an avanos medical inc.Product is defective or caused serious injury.
|
|
Manufacturer Narrative
|
Correction: g3 the initial report alert date is 14 dec 2021, not 15 dec 2021 as was reported previously.One used ng tube was returned for evaluation.The distal portion of the tube was noted to be missing.The broken off end was not received with the sample.The broken edge of the tube was evaluated under magnification and found to be smooth to slightly rough in appearance.The complaint is confirmed as reported however, no root cause could be conclusively determined.All information reasonably known as of 21 mar 2022 has been included in this health authority report.Should additional information be obtained, a follow-up health authority report will be provided.The information provided by avanos medical inc.Represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to avanos medical inc.Avanos medical inc.Has no independent knowledge of the event reported but is relaying the information that was provided by the user facility where the incident occurred.This product incident is documented in the avanos medical complaint database and identified as complaint (b)(4).This information is submitted pursuant to 21cfr803, in compliance with the medical device reporting requirement and should not be considered to be an admission that an avanos medical inc.Product is defective or caused serious injury.
|
|