• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEPUY CMW - 9610921 SMARTSET GMV 40G US EO; BONE CEMENT : BONE CEMENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY CMW - 9610921 SMARTSET GMV 40G US EO; BONE CEMENT : BONE CEMENT Back to Search Results
Model Number 5450-50-501
Device Problem Loss of or Failure to Bond (1068)
Patient Problems Scar Tissue (2060); Joint Laxity (4526); Unspecified Tissue Injury (4559)
Event Date 12/04/2020
Event Type  Injury  
Manufacturer Narrative
(b)(4).Initial reporter occupation: lawyer.Dmf# (b)(4); trade name gentamicin sulphate; active ingredient(s) gentamicin sulphate; dosage form - powder; strength 1.0g active in our cements.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event Description
Patient received a left attune knee to treat severe osteoarthritis.The patella was resurfaced and depuy cement x 2 was utilized.There were no indicated intra-operative complications.Patient received a left knee revision to address tibial tray loosening at the cement to implant interface, severe tibia bone loss, tibial tray migration, removal of dense scar tissue, instability, and distal and femoral bone somewhat compromised from significant osteolysis particularly the posterior condyles.The tibial tray, tibial insert, and femoral component were revised.After removal of heterotopic bone and dense scar tissue, the well-fixed patella component tracked well and was retained.The patient was revised with competitor products and cement.There were no indicated intra-operative complications.Doi: (b)(6) 2015, dor: (b)(6) 2020, left knee.
 
Manufacturer Narrative
Product complaint # (b)(4).Investigation summary: no device associated with this report was received for examination.The information received will be retained for potential series investigations if triggered by trend analysis, post market surveillance, or other events within the quality system.Depuy considers the investigation closed.Should additional information be received, the information will be reviewed and the investigation will be re-opened as necessary.Device history lot: a previous dhr review was conducted for (b)(4) against the provided smartset gmv 40g us eo product code 545050501, lot 8068693 combination.A device history record (dhr) review per (b)(4) found two unrelated non conformances on this batch.Micro and sterility tests passed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SMARTSET GMV 40G US EO
Type of Device
BONE CEMENT : BONE CEMENT
Manufacturer (Section D)
DEPUY CMW - 9610921
cornford rd
blackpool FY4 4 QQ
UK  FY4 4QQ
Manufacturer (Section G)
DEPUY CMW - 9610921
cornford rd
blackpool FY4 4 QQ
UK   FY4 4QQ
Manufacturer Contact
kara ditty-bovard
700 orthopaedic dr.
warsaw, IN 46581-0988
6107428552
MDR Report Key13109574
MDR Text Key282904514
Report Number1818910-2021-28933
Device Sequence Number1
Product Code MBB
UDI-Device Identifier10603295174295
UDI-Public10603295174295
Combination Product (y/n)Y
Reporter Country CodeUS
PMA/PMN Number
K081163
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Consumer
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 12/29/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/29/2021
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date03/30/2017
Device Model Number5450-50-501
Device Catalogue Number545050501
Device Lot Number8068693
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Date Manufacturer Received01/24/2022
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured07/02/2015
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
ATTUNE FB TIB BASE SZ 5 CEM; ATTUNE MEDIAL DOME PAT 38MM; ATTUNE PS FB INSRT SZ 5 7MM; ATTUNE PS FEM LT SZ 5 CEM; SMARTSET GMV 40G US EO; SMARTSET GMV 40G US EO
Patient Outcome(s) Required Intervention;
Patient Age73 YR
Patient SexFemale
-
-