According to the reporter, pre-operatively, due to jaw malformation, the device had a jamming situation.A new product was used to resolve the issue.There was no patient involvement.
|
Additional information: g3, h3, h6 h3 evaluation summary: medtronic conducted an investigation based upon all information received.The device was available for evaluation.Visual inspection noted the jaws of the instrument were out of alignment.Functionally, the instrument was first test fired once in air so that the clip formation could be observed.The clip displayed a scissor formation.Four clips were fired on test media with scissor formation.The jaw and handle moved smoothly through the firing cycle and returned to the open position.Each remaining clip loaded properly in the jaw.When the cartridge was empty, the interlock engaged and prevented the jaws from approximating.It was reported that the jaws of the device were broken, and the device jammed.The reported issues were confirmed.The product analysis noted evidence that the device was not used as intended.This issue may occur if the distal end of the device is subjected to excessive manipulation during application of the instrument.The manufacturing records for each device are thoroughly reviewed prior to release to ensure that it meets all medtronic quality specifications.The instructions included with this device provide the following guidance: do not excessively twist or torque the jaws.Excessive twisting or torqueing the jaws may dislodge clip from the jaws or cause clip malformation after jaw closure.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|