This report is being supplemented to provide additional information based on the legal manufacturer's investigation.The device history record (dhr) was unable to be reviewed for this device since the serial number was not provided.However, olympus only releases products to market that meet all manufacturing specifications and final product release criteria.The case is not due to design.Based on the results of the legal manufacturer's investigation, the root cause of the reported device not being brushed during reprocessing could not be identified.However, it is the possible the cause was related to the facility not having the needed brush on hand.Per the legal manufacturer's instructions for use (listed below), the phenomenon might have been prevented: 7.1 required reprocessing equipment.Preparation of the equipment.Prior to cleaning and disinfection or sterilization, prepare the equipment shown in figure 7.1.Olympus will continue to monitor field performance for this device.
|