The mayfield skull clamp (a1059) was returned for evaluation: failure analysis - the investigation of the returned device showed that the reported complaint was confirmed from the evaluation.On inspection it was noted that the device has rotational and lateral movement, but a residue buildup was present internally.New components were added to replace the visibly worn internal parts.General maintenance and cleaning was performed as well.Root cause - unit was received in used condition with visible residue buildup and worn internal parts likely due to excessive use over time/environmental damage.This unit is beyond integra¿s 7 years recommended life cycle (manufactured 2007).No further investigation is required based on risk acceptability and no manufacturing defects were noted.This will be monitored and trended going forward.
|