• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION WOLVERINE CORONARY CUTTING BALLOON MONORAIL; CATHETER, PERCUTANEOUS TRANSLUMINAL CORONARY ANGIOPLASTY (PTCA), CUTTING/SCORING

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION WOLVERINE CORONARY CUTTING BALLOON MONORAIL; CATHETER, PERCUTANEOUS TRANSLUMINAL CORONARY ANGIOPLASTY (PTCA), CUTTING/SCORING Back to Search Results
Model Number 3851
Device Problem Material Rupture (1546)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 12/09/2021
Event Type  malfunction  
Event Description
It was reported that the balloon ruptured.The target lesion was located in the moderately tortuous and moderately calcified left anterior descending artery.A 10mmx2.50mm wolverine coronary cutting balloon was selected for use.During the procedure, at tenth inflation, it was noted that the balloon ruptured at 10atm within 5 seconds.The device was removed using normal method without problem.The procedure was completed with another of the same device.No patient complications were reported and the patient status was good.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
WOLVERINE CORONARY CUTTING BALLOON MONORAIL
Type of Device
CATHETER, PERCUTANEOUS TRANSLUMINAL CORONARY ANGIOPLASTY (PTCA), CUTTING/SCORING
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
two scimed place
maple grove MN 55311
Manufacturer (Section G)
BOSTON SCIENTIFIC IRELAND LIMITED
ballybrit business park
galway
EI  
Manufacturer Contact
jay johnson
two scimed place
maple grove, MN 55311
7634942574
MDR Report Key13120390
MDR Text Key283152309
Report Number2134265-2021-16363
Device Sequence Number1
Product Code NWX
Combination Product (y/n)N
Reporter Country CodeJA
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative,Distributor
Reporter Occupation Physician
Type of Report Initial
Report Date 12/30/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/30/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date12/19/2021
Device Model Number3851
Device Catalogue Number3851
Device Lot Number0024962395
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received12/16/2021
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured12/20/2019
Is the Device Single Use? Yes
Type of Device Usage Initial
Patient Sequence Number1
-
-