• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: TERUMO BCT SPECTRA OPTIA; SPECTRA OPTIA COLLECT SET

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

TERUMO BCT SPECTRA OPTIA; SPECTRA OPTIA COLLECT SET Back to Search Results
Model Number 12120
Device Problem Adverse Event Without Identified Device or Use Problem (2993)
Patient Problems Dyspnea (1816); Cramp(s) /Muscle Spasm(s) (4521)
Event Date 06/16/2021
Event Type  Injury  
Event Description
The customer reported that a donor had severe tetany with dyspnoea during a spectra optia procedure.Per the customer the patient is recovered.The collection set is not available for return because it was discarded by the customer.
 
Manufacturer Narrative
Lot number and expiry information are not available at this time.Investigation is in process.A follow-up report will be provided.
 
Event Description
The customer reported that a donor had severe tetany with dyspnoea during a spectra optia procedure.Per the customer the patient is recovered.The collection set is not available for return because it was discarded by the customer.
 
Manufacturer Narrative
This report is being filed to provide additional information in b.6 and h.10.Lot number and expiry information are not available at this time.Investigation: the run data file (rdf) was analyzed for this event.Review of the run data file did not indicate a root cause for the reported citrate reaction of the patient.The system operated as intended and the procedures were run within standard operating limits (i.E., not in ¿caution status¿).There were no signals or alarms in the dlog that would indicate any issues in the run, specifically the amount of ac delivered.The commanded ac pump speed and the measured ac pump speed were accurate and did not show any discrepancy in pump volume.Investigation is in process.A follow-up report will be provided.
 
Manufacturer Narrative
This report is being filed to provide additional information in h.6 and h.10.Investigation: the run data file (rdf) was analyzed for this event.Review of the run data file did not indicate a root cause for the reported citrate reaction of the patient.The system operated as intended and the procedures were run within standard operating limits (i.E., not in ¿caution status¿).There were no signals or alarms in the dlog that would indicate any issues in the run, specifically the amount of ac delivered.The commanded ac pump speed and the measured ac pump speed were accurate and did not show any discrepancy in pump volume.The lot number was not provided, therefore, a dhr search could not be conducted for this specific incident.All lots must meet acceptance criteria for release.According to therapeutic apheresis: a physician's handbook, adverse events occur during therapeutic procedures with a frequency of 4.8%.Some of the most common reactions include fever, urticaria, hypocalcemic symptoms, pruritus, dyspnea, tachycardia, and mild hypotension.Transient hypocalcemia associated with apheresis is usually well tolerated.Symptoms often show as parasethesia (tingling) but patients may also experience unusual taste, nausea, lightheadedness, shivering, and tremors.Severe hypocalcemia may also cause muscle contractions and can progress to tetany and seizures if hypocalcemia escalates and is not corrected.Root cause: review of the run data file did not indicate a root cause for the reported citrate reaction of the donor.The system operated as intended and the procedures were run within standard operating limits (i.E., not in ¿caution status¿).There were no signals or alarms in the dlog that would indicate any issues in the run, specifically the amount of ac delivered.The commanded ac pump speed and the measured ac pump speed were accurate and did not show any discrepancy in pump volume.A root cause assessment was performed for this complaint.Based on the available information a definitive root cause could not be determined.Citrate reactions occur due to decreased ionized calcium in circulation as a result of exogenous citrate administered during the apheresis procedure and are influenced by donor physiology, the rate of ac infusion, and/or the length of the procedure.These symptoms may be treated with oral or intravenous calcium supplements or by adjusting the ac infusion rate.
 
Event Description
The customer reported that a donor had severe tetany with dyspnoea during a spectra optia procedure.Per the customer the patient is recovered.The collection set is not available for return because it was discarded by the customer.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SPECTRA OPTIA
Type of Device
SPECTRA OPTIA COLLECT SET
Manufacturer (Section D)
TERUMO BCT
lakewood CO 80215
Manufacturer (Section G)
TERUMO BCT
10810 w. collins ave
lakewood CO 80215
Manufacturer Contact
scot hilden
10810 w. collins ave
lakewood, CO 80215
MDR Report Key13142493
MDR Text Key286664997
Report Number1722028-2022-00001
Device Sequence Number1
Product Code GKT
UDI-Device Identifier05020583121201
UDI-Public05020583121201
Combination Product (y/n)N
Reporter Country CodeGM
PMA/PMN Number
BK150251
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional
Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date 01/03/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received01/03/2022
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Model Number12120
Device Catalogue Number12120
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received09/20/2022
Was Device Evaluated by Manufacturer? No
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
ACD-A; ACD-A; ACD-A
Patient Outcome(s) Other;
Patient Age28 YR
Patient SexMale
Patient Weight64 KG
-
-