Model Number 862116 |
Device Problem
Display or Visual Feedback Problem (1184)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 12/08/2021 |
Event Type
malfunction
|
Manufacturer Narrative
|
A follow-up report will be submitted upon completion of the investigation.
|
|
Event Description
|
It was reported the message on the display freezes.It is unknown if the device was in clinical use at the time of the event, no adverse event or patient harm was reported.
|
|
Manufacturer Narrative
|
A philips field service engineer (fse) went to the customer site.The fse determined that the issue was associated with a defective nurse call cable.After parts replacement, the device was returned to functional use with no further issues identified.
|
|
Event Description
|
It was reported the message on the display freezes.The device was reported to be in use on a patient, but no adverse event to the patient or user was reported.
|
|