Model Number N/A |
Device Problems
Failure to Cut (2587); Pressure Problem (3012)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 12/01/2021 |
Event Type
malfunction
|
Manufacturer Narrative
|
This complaint is recorded by zimmer biomet under (b)(4).A follow up/ final report will be submitted once investigation is complete.
|
|
Event Description
|
It was reported that the device did not receive enough compressed air.For this reason the device was not working properly and skin grafts become ragged.There is no harm or delay reported.No adverse events were reported as a result of this malfunction.
|
|
Manufacturer Narrative
|
This event has been recorded by zimmer biomet under (b)(4).Disregard h5 and a5 as they were previously reported on the initial mdr.Review of the most recent repair record determined the motor speed was unstable and the motor pin was damaged.The motor and pin were replaced and resolved the reported issue.A definitive root cause cannot be determined.The event is confirmed.If any further information is found which would change or alter any conclusions or information, a supplemental report will be filed accordingly.Zimmer biomet will continue to monitor for trends.
|
|
Event Description
|
There was no additional information.
|
|