• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ZIMMER SURGICAL, INC. ZIMMER AIR DERMATOME; DERMATOME, PNEUMATICALLY-POWERED

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ZIMMER SURGICAL, INC. ZIMMER AIR DERMATOME; DERMATOME, PNEUMATICALLY-POWERED Back to Search Results
Model Number N/A
Device Problems Failure to Cut (2587); Pressure Problem (3012)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 12/01/2021
Event Type  malfunction  
Manufacturer Narrative
This complaint is recorded by zimmer biomet under (b)(4).A follow up/ final report will be submitted once investigation is complete.
 
Event Description
It was reported that the device did not receive enough compressed air.For this reason the device was not working properly and skin grafts become ragged.There is no harm or delay reported.No adverse events were reported as a result of this malfunction.
 
Manufacturer Narrative
This event has been recorded by zimmer biomet under (b)(4).Disregard h5 and a5 as they were previously reported on the initial mdr.Review of the most recent repair record determined the motor speed was unstable and the motor pin was damaged.The motor and pin were replaced and resolved the reported issue.A definitive root cause cannot be determined.The event is confirmed.If any further information is found which would change or alter any conclusions or information, a supplemental report will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Event Description
There was no additional information.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ZIMMER AIR DERMATOME
Type of Device
DERMATOME, PNEUMATICALLY-POWERED
Manufacturer (Section D)
ZIMMER SURGICAL, INC.
200 west ohio avenue
dover OH 44622
Manufacturer (Section G)
ZIMMER SURGICAL, INC.
200 west ohio avenue
dover OH 44622
Manufacturer Contact
christina arnt
56 e. bell drive
warsaw, IN 46582
5745273773
MDR Report Key13160558
MDR Text Key283480143
Report Number0001526350-2021-01338
Device Sequence Number1
Product Code GFD
Combination Product (y/n)N
Reporter Country CodeFI
PMA/PMN Number
EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,User Facility
Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date 03/03/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received01/04/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberN/A
Device Catalogue Number00880100100
Device Lot Number63130445
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/21/2021
Was the Report Sent to FDA? No
Date Manufacturer Received02/21/2022
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured08/03/2015
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Removal/Correction NumberN/A
Patient Sequence Number1
-
-