SMITH & NEPHEW, INC. GII UNIVERSAL EXTRACTOR; PROSTHESIS, KNEE, FEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, METAL/POLYMER Back to Search Results
Model Number 71440366
Device Problems Break (1069); Shipping Damage or Problem (1570)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 12/16/2021
Event Type  malfunction  
Manufacturer Narrative
(b)(4).
 
Event Description
It was reported that a genesis ii universal extractor broke in shipping.As this was noticed upon field inspection, there was no patient involvement.
 
Manufacturer Narrative
H3, h6: the associated device was returned and evaluated.The manufacturing process of the device did not contribute to the reported event.A visual inspection confirmed one end of the device is fractured, rendering it inoperative.A review of complaint history did not reveal similar events for the listed device.No adverse trend was found for this event.At this time, we do not have reason to suspect that the product failed to meet any product specifications at the time of manufacture.A review of the risk management file revealed this failure mode was previously identified.The anticipated risk level is still adequate.Assessment of historical escalated cases concluded that there are no prior actions related to this device and failure mode.This device is a reusable instrument that can be exposed to numerous surgeries.Damage from prolonged use, misuse or rough handling are probable causes of the reported event.We recommend that all reusable instruments be routinely inspected for wear and damage and replaced as necessary.Based on this investigation, the need for corrective action is not indicated.Should additional information be received, the complaint will be reopened.No further investigation is warranted for this complaint; however, we will continue to monitor future complaints and investigate as necessary.We consider this investigation closed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
GII UNIVERSAL EXTRACTOR
Type of Device
PROSTHESIS, KNEE, FEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, METAL/POLYMER
Manufacturer (Section D)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
Manufacturer (Section G)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
Manufacturer Contact
holly topping
7000 west william cannon drive
austin
austin, TX 78735
5123913905
MDR Report Key13160961
MDR Text Key283258726
Report Number1020279-2022-00065
Device Sequence Number1
Product Code HRY
UDI-Device Identifier03596010215192
UDI-Public03596010215192
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K121393
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Company Representative
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 01/31/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received01/04/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number71440366
Device Catalogue Number71440366
Device Lot NumberUNKNOWN
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/28/2021
Is the Reporter a Health Professional? No
Date Manufacturer Received01/28/2022
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-