Model Number CI-1600-05 |
Device Problems
Mechanical Problem (1384); Adverse Event Without Identified Device or Use Problem (2993)
|
Patient Problems
Headache (1880); Nausea (1970); Tinnitus (2103); Dizziness (2194)
|
Event Type
malfunction
|
Event Description
|
The recipient is reportedly experiencing headache, dizziness, tinnitus and nausea with and without device use.Programming adjustments have been made, however the issues did not resolve.A review of the recipient's test data indicated impedance issues.Revision surgery is scheduled.
|
|
Event Description
|
The recipient is reportedly experiencing headache, dizziness, tinnitus and nausea with and without device use.Programming adjustments have been made, however the issues did not resolve.A review of the recipient's test data indicated impedance issues.Revision surgery is scheduled.
|
|
Manufacturer Narrative
|
The recipient's device was explanted.The recipient was reimplanted with another cochlear device.Disclaimer: advanced bionics does not intend that this report be any admission of liability, fault or product defect.
|
|
Manufacturer Narrative
|
Advanced bionics considers the investigation into this reportable event as closed.The external visual inspection revealed torn silicone overmold on the top cover, as well as a severed electrode.These are believed to have occurred during revision surgery.The device passed the photographic imaging inspection.System lock was verified.The electrode condition prevented an electrical test from being performed.The device passed some of the electrical tests performed.The device passed the mechanical test performed.The failure of this device is attributed to an electrode short in the electrode pocket.A corrective action was implemented.This version of the ultra is no longer distributed.This is the final report.Disclaimer: advanced bionics does not intend that this report be any admission of liability, fault or product defect.
|
|