RESPIRONICS CALIFORNIA, LLC RESPIRONICS; VENTILATOR, CONTINUOUS, MINIMAL VENTILATORY SUPPORT, FACILITY USE
|
Back to Search Results |
|
Model Number V60 |
Device Problem
Increase in Pressure (1491)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 12/15/2021 |
Event Type
malfunction
|
Event Description
|
It was reported to philips by a customer that the unit's high 02 pressure 02 was not available.Based upon the information provided, the being set up for use at the time of the event reported.No harm or injury has been indicated or alleged.The device was evaluated remotely by a philips remote service engineer (rse).Upon further inspection and review of the device, the customer stated that the unit alarms "high oxygen pressure".The rse determined that the unit¿s oxygen inlet pressure display showed 113 psi even after removing the oxygen source.The rse advised the customer that the da pcba may be faulty.The customer had a third-party vendor replace the data acquisition pcba to resolve the reported issue.
|
|
Search Alerts/Recalls
|
|
|