Model Number FB-15RBS
Device Problem Fracture (1260)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 12/07/2021
Event Type  malfunction  
Event Description
The root of ift was easy to crack.The time of event is unknown.There was no report of patient harm.
 
Manufacturer Narrative
This device is not distributed in the usa, therefore the pma/510(k) number is not applicable.If additional information becomes available, a supplemental report will be filed with the new information.
 
Manufacturer Narrative
Evaluation summary: the root of ift was broken, because of man-made force.The scope was repaired by the third party and returned to the hospital.Correction information: h6: coding changed, based on the investigation result.Additional information: d8: this device was serviced by a third party.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PENTAX
Type of Device
FIBEROPTICBRONCHOSCOPE 2.0C 4.8TP RO PLS
Manufacturer (Section D)
HOYA CORPORATION PENTAX TOKYO OFFICE
tsutsujigaoka 1-1-110
akishima-shi, tokyo 196-0 012
JA  196-0012
Manufacturer (Section G)
HOYA CORPORATION PENTAX TOKYO OFFICE
tsutsujigaoka 1-1-110
akishima-shi, tokyo 196-0 012
JA   196-0012
Manufacturer Contact
william goeller
3 paragon drive
montvale, NJ 07645
8004315880
MDR Report Key13173183
MDR Text Key289218675
Report Number9610877-2022-00069
Device Sequence Number1
Product Code EOQ
Combination Product (y/n)N
Reporter Country CodeCH
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 05/12/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received01/05/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberFB-15RBS
Is the Reporter a Health Professional? Yes
Date Manufacturer Received12/22/2021
Was Device Evaluated by Manufacturer? No
Date Device Manufactured09/18/2018
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-