As reported, "something came out of the scope while it was in the patient.Customer stated no one could verify as the user did not show what came out from the scope.Customer stated ," they are not sure about the insertion tube, it might need to be repaired or replaced".The issue found during an unspecified procedure.There was no patient harm or injury reported due to the event.No user injury was reported.
|
This report is being supplemented to provide additional information based on the legal manufacturer's final investigation.A review of the device history record found no deviations that could have caused or contributed to the reported issue.It has been over 9 years since the subject device was manufactured.Based on the results of the investigation, it is likely that because of inappropriate reprocessing after the previous procedure, foreign materials entered into the air/water channel, and then the materials were pushed by water or air during procedure.A definitive root cause cannot be identified.Users can detect the suggested event properly by handling the device in accordance with the following instructions for use (ifu)."·inspection of the endoscope : inspect the external surface of the entire insertion section, including the bending section and the distal end for dents, bulges, swelling, scratches, peeling of coating, holes, sagging, transformation, bends, adhesion of foreign bodies, missing parts, protruding objects, or other irregularities.·inspection of the endoscopic system, - inspection of the air-feeding function, - inspection of the objective lens cleaning function.·inspection of ancillary equipment - inspect the following equipment as described in their respective instruction manuals." users can decrease/prevent the suggested event by handling the device in accordance with the following ifu."·an insufficiently reprocessed endoscope and/or accessories may pose an infection control risk to the patients and/or operators who contact them.If the endoscope is not immediately cleaned after each procedure, residual organic debris will begin to solidify and it may be difficult to effectively reprocess the endoscope.·to prevent clogging of the air/water nozzle of the endoscope, flush water into the air channel of the endoscope, using the aw channel cleaning adapter (mh-948) after each patient procedure.·troubleshooting - the accessories are consumables.Olympus does not repair accessory parts.If an accessory part becomes damaged, contact olympus to purchase a replacement" olympus will continue to monitor the field performance of this device.
|