Model Number N/A |
Device Problem
Output Problem (3005)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 12/17/2021 |
Event Type
malfunction
|
Event Description
|
It was reported that the device didn't produce consistent depth, it failed to produce a uniform depth in skin graft.The event occurred during surgery.There was no harm or delay.No adverse event was reported as a result of this malfunction.
|
|
Manufacturer Narrative
|
This event has been recorded by zimmer biomet under (b)(4).Once the investigation is completed, a follow-up/final report will be submitted.
|
|
Manufacturer Narrative
|
This complaint has been recorded under (b)(4).Review of the most recent repair record determined the device was out of calibration and recalibrated and resolved the reported issue.Review of the previous repair record identified unrelated repairs to the reported event.The device was tested and found to be functioning to specifications prior to release to the customer.Device is used for treatment.The event is confirmed.
|
|
Event Description
|
No additional information available.
|
|