• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ZIMMER SURGICAL, INC. ZIMMER AIR DERMATOME

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ZIMMER SURGICAL, INC. ZIMMER AIR DERMATOME Back to Search Results
Model Number N/A
Device Problem Output Problem (3005)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 12/17/2021
Event Type  malfunction  
Event Description
It was reported that the device didn't produce consistent depth, it failed to produce a uniform depth in skin graft.The event occurred during surgery.There was no harm or delay.No adverse event was reported as a result of this malfunction.
 
Manufacturer Narrative
This event has been recorded by zimmer biomet under (b)(4).Once the investigation is completed, a follow-up/final report will be submitted.
 
Manufacturer Narrative
This complaint has been recorded under (b)(4).Review of the most recent repair record determined the device was out of calibration and recalibrated and resolved the reported issue.Review of the previous repair record identified unrelated repairs to the reported event.The device was tested and found to be functioning to specifications prior to release to the customer.Device is used for treatment.The event is confirmed.
 
Event Description
No additional information available.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ZIMMER AIR DERMATOME
Type of Device
DERMATOME
Manufacturer (Section D)
ZIMMER SURGICAL, INC.
200 west ohio avenue
dover OH 44622
Manufacturer (Section G)
ZIMMER SURGICAL, INC.
200 west ohio avenue
dover OH 44622
Manufacturer Contact
christina arnt
56 e. bell drive
warsaw, IN 46582
5745273773
MDR Report Key13175591
MDR Text Key283305628
Report Number0001526350-2022-00014
Device Sequence Number1
Product Code GFD
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type User Facility
Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date 02/17/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received01/06/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberN/A
Device Catalogue Number00880100100
Device Lot Number64276928
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/11/2022
Was the Report Sent to FDA? No
Date Manufacturer Received02/17/2022
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured01/15/2019
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Removal/Correction NumberN/A
Patient Sequence Number1
-
-