|
Model Number INT30018UX |
Device Problems
Migration or Expulsion of Device (1395); Device Dislodged or Dislocated (2923); Material Deformation (2976)
|
Patient Problem
Insufficient Information (4580)
|
Event Date 12/09/2021 |
Event Type
Injury
|
Manufacturer Narrative
|
If information is provided in the future, a supplemental report will be issued.
|
|
Event Description
|
An attempt was made to use an integrity rx coronary bare metal stent to treat the left vertebral artery.The device was inspected with no issues noted.The device did not pass through a previously-deployed stent.Excessive force was not used during delivery.It was reported that stent dislodgement occurred following a failed delivery.It was stated that the integrity bare metal stent was placed in a guide sheath but the guide sheath became dislodged, therefore the stent was removed to reposition in the guide sheath.When an attempt was made to use a balloon to deploy the stent, it was noticed that the stent had fallen off the balloon and injection showed that the stent had migrated and was floating in the left vertebral artery unexpanded.The stent was retrieved and removed possibly using a snare and the procedure was continued.No further patient injury was reported.
|
|
Manufacturer Narrative
|
Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.
|
|
Manufacturer Narrative
|
Product analysis summary: the dislodged stent returned caught in a snare.The delivery system did not return for analysis.Deformation was evident to the dislodged stent.Additional information: lot number provided.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
|
Search Alerts/Recalls
|
|
|