• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITH & NEPHEW, INC. SET MENISCUS MENDER II DISPOSABLE; PASSER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITH & NEPHEW, INC. SET MENISCUS MENDER II DISPOSABLE; PASSER Back to Search Results
Model Number 7209485
Device Problems Break (1069); Device Damaged Prior to Use (2284)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 12/17/2021
Event Type  malfunction  
Event Description
It was reported that during use in a meniscus repair procedure, the meniscus mender was found to be already broken when the package was opened.The procedure was successfully completed without delay using a smith and nephew back up device.No patient complications were reported.
 
Manufacturer Narrative
Internal complaint reference: case-(b)(4).
 
Manufacturer Narrative
Internal complaint reference (b)(4).H10: h3, h6: the reported device was received for evaluation.It was determined the device contributed to the reported event.A complaint history review concluded this was a repeat issue.A review of device records showed there were no indications to suggest that the product did not meet manufacturing specification upon release for distribution.A review of risk management found that the anticipated risk level is no longer adequate.A review of the device drawings found a stress relief collar is now attached at the connection between the shaft and the hub of the suture capture loop.This design change was implemented as a result of a corrective action initiated to address the reported issue.A visual inspection of the returned device found that it is not in its original packaging.Only one meniscal suture loop was returned, and it is separated between the shaft and hub.There is no debris on the device.The complaint was confirmed, and the root cause was associated with device design.Please refer to the instructions for use for recommendations on proper use of the device and potential troubleshooting methods to prevent future reoccurrence of the reported event.A corrective action for this failure mode is in place.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SET MENISCUS MENDER II DISPOSABLE
Type of Device
PASSER
Manufacturer (Section D)
SMITH & NEPHEW, INC.
130 forbes blvd.
mansfield MA 02048
Manufacturer (Section G)
SMITH & NEPHEW, INC.
130 forbes blvd.
mansfield MA 02048
Manufacturer Contact
holly topping
7000 west william cannon drive
austin, TX 78735
5123913905
MDR Report Key13193367
MDR Text Key283420600
Report Number1219602-2022-00045
Device Sequence Number1
Product Code HWQ
UDI-Device Identifier03596010468482
UDI-Public03596010468482
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
K885311
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign,Consumer,Health Professional,User Facility,Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 03/08/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received01/07/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number7209485
Device Catalogue Number7209485
Device Lot Number2055687
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/18/2022
Is the Reporter a Health Professional? Yes
Date Manufacturer Received03/07/2022
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured04/27/2020
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-