RESPIRONICS CALIFORNIA, LLC RESPIRONICS; VENTILATOR, CONTINUOUS, MINIMAL VENTILATORY SUPPORT, FACILITY USE
|
Back to Search Results |
|
Model Number V60 VENT, CHINA OPT: CFLEX, AVAPS |
Device Problems
Volume Accuracy Problem (1675); Noise, Audible (3273)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 12/19/2021 |
Event Type
malfunction
|
Event Description
|
At 14:43 on (b)(6) 2021, when the patient was using the ventilator, the nurse found that the tidal volume of the ventilator could not be reached all the time and found that there was an abnormal sound inside the ventilator.She immediately replaced the ventilator and reported to the equipment department for repair.After the engineer arrived at the site, he found that the blower in the pneumatic assembly made abnormal noise and was covered with dust.The engineer cleaned the dust as well as replaced the blower and filter.The device passed the self-inspection.
|
|