Model Number OER-AW
Device Problems Leak/Splash (1354); Material Deformation (2976)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Type  malfunction  
Event Description
As reported, leakage in the inner tube was observed.The issue found at preparation for use.There is no patient involvement associated on this reported event.Device evaluation found alcohol tank lid damage.This report is being submitted for alcohol lid damage.
 
Manufacturer Narrative
The subject device was received and evaluated at service repair olympus korea.Evaluation found the customer reported issue was confirmed.Device evaluation the following findings were found : channel block leakage.3 ports valve connector leakage.Corrosion due to dc pump leakage.Alcohol tank lid damage tube deformation.Top cover packing deformation.Investigation is ongoing.This report will be supplemented accordingly following investigation.
 
Manufacturer Narrative
This report is being supplemented to provide additional information based on the legal manufacturer's final investigation and correction to g2.G2 - checked "other" to add the country south korea.A review of the device history record found no deviations that could have caused or contributed to the reported issue.It has been over 11 years since the subject device was manufactured.Based on the results of the investigation, the cause of the alcohol tank lid damage likely occurred from excessive stress being applied to the connector by coming in contact with something hard or from aged deterioration.The following information is stated in the instructions for use: ¿check the following for each connector.The connector should be fixed firmly.The o-rings should be free of abnormalities such as cracks, tears, or dents.¿.Olympus will continue to monitor field performance for this device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
OER-AW 110V
Type of Device
ENDOSCOPE REPROCESSOR
Manufacturer (Section D)
OLYMPUS MEDICAL SYSTEMS CORP.
2951 ishikawa-cho
hachioji-shi, tokyo-to 192-8 507
JA  192-8507
Manufacturer Contact
kazutaka matsumoto
2951 ishikawa-cho
hachioji-shi, tokyo-to 192-8-507
JA   192-8507
426425177
MDR Report Key13196912
MDR Text Key288666528
Report Number8010047-2022-01101
Device Sequence Number1
Product Code FEB
Combination Product (y/n)N
Reporter Country CodeKS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign,Health Professional,Company Representative
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 02/08/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received01/07/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberOER-AW
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/14/2021
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received02/01/2022
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured08/24/2010
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-