MAKO SURGICAL CORP. MCK TIBIAL BASEPLATE-LM/RL-SZ 5; PROSTHESIS, KNEE PATELLOFEMOROTIBIAL, PARTIAL, SEMI-CONSTRAINED, CEMENTED, POLYM
|
Back to Search Results |
|
Model Number 180605 |
Device Problem
Fracture (1260)
|
Patient Problem
Ambulation Difficulties (2544)
|
Event Date 12/14/2021 |
Event Type
Injury
|
Manufacturer Narrative
|
As part of normal complaint follow-up, an evaluation of the event has been initiated by mako surgical.A supplemental report will be submitted when additional information becomes available.Device not returned.
|
|
Event Description
|
Tibia baseplate broke in half.Explanted.
|
|
Manufacturer Narrative
|
Reported event: an event regarding crack/fracture involving a mako baseplate was reported.The event was not confirmed.Method & results: device evaluation and results: not performed as product was not returned.Clinician review: no medical records were received for review with a clinical consultant.Device history review: review of the device history records indicate the devices were manufactured and accepted into final stock with no relevant reported discrepancies.Complaint history review: there have been no other similar events for the lot referenced.Conclusion: the event could not be confirmed as insufficient information was provided.Further information such as return of the device, pre- and post-operative x-rays, primary operative report as well as patient history and follow-up notes are needed to complete the investigation for determining root cause.No further investigation for this event is possible at this time.If additional information become available to indicate further evaluation is warranted, this record will be reopened.
|
|
Event Description
|
Tibia baseplate broke in half.Explanted.
|
|