Model Number 242436 |
Device Problems
Display or Visual Feedback Problem (1184); Difficult to Open or Close (2921)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 01/06/2022 |
Event Type
malfunction
|
Manufacturer Narrative
|
Udi: (01)10886705028634(21)b001111.To date the device has not been returned.If the device or further details are received at a later date a supplemental medwatch will be sent.
|
|
Event Description
|
It was reported by the sales rep that during an unknown surgery on (b)(6) 2022, it was observed that the coupler ac zoom device was gritty in nature where the coupler was having trouble closing and also was consistently fogging due to poor seal on the scope.There were no adverse patient consequences nor surgical delay reported.No additional information was provided.
|
|
Manufacturer Narrative
|
This report is being submitted in pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by mitek or its employees that the report constitutes an admission that the device, mitek, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.H6: the actual device has been returned and is currently pending evaluation.Once the device has been evaluated, a supplemental medwatch report will be sent accordingly.
|
|
Manufacturer Narrative
|
This report is being submitted in pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by mitek or its employees that the report constitutes an admission that the device, mitek, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.H10 additional narrative: investigation summary:the complaint device was received at the manufacturing site and evaluated.Per service reports, this complaint cannot be confirmed.During the service evaluation the following defects were identified: damaged grasping mechanism.External damage of adjusting rings.Minor scratches on the unit.The endoscope grasping mechanism and adjusting rings including o-ring seals was replaced to resolve the issues.After repair, the device was found to be working according to the specifications.The user error was identified as the root cause for the device failure during the service evaluation.Manufacturing record evaluation is not required as the reported event is not associated with the manufacturing process and/or the potential cause of the defect cannot be associated to manufacturing.At this point in time, no corrective action is required, and no further action is warranted.Depuy mitek will continue to track any related complaints within this device family as a means of monitoring the extent with which this complaint is observed in the field.
|
|