This report is being supplemented to provide additional information based on the legal manufacturer's final investigation.A review of the device history record found no deviations that could have caused or contributed to the reported issue.It has been over 5 years since the subject device was manufactured.Based on the results of the investigation, it was confirmed by the olympus field service engineer that too much force was applied to drawer by the user.The following information is stated in the instructions for use: "the oer-pro requires routine maintenance and inspection.In addition to checks before use, the person in charge of maintenance and administration of the medical equipment at the hospital should periodically check all of the items described in this manual.If any irregularity is observed, do not use the equipment and inspect it as described in 8.1, ¿troubleshooting guide¿ on page 274.If the irregularity is still present, the equipment must be repaired prior to next use." olympus will continue to monitor field performance for this device.
|