Model Number A3059
Device Problems Mechanical Problem (1384); Noise, Audible (3273)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Type  malfunction  
Manufacturer Narrative
The mayfield composite series skull clamp (a3059) was returned for evaluation: device history record (dhr): the dhr was reviewed and shows no abnormalities related to the reported failure.Failure analysis - the investigation of the returned device showed that the reported complaint was confirmed from the evaluation.The unit received had a sticking index knob and the teeth were not meshing.The lock ratchet was replaced due to worn teeth and general maintenance and cleaning were performed.Root cause - complaint is confirmed via inspection of the unit.Unit received had a sticking index knob and the teeth were not meshing.The lock ratchet was replaced due to worn teeth.No further investigation required based on the acceptability of risk and no adverse trends identified.This will be monitored and trended going forward.
 
Event Description
A facility reported that the black locking knob on the mayfield composite skull clamp (a3059) makes a squeaking sound when being locked and unlocked.It is also difficult to turn black locking knob into the locked position.It is unknown if there was patient involvement; however, no patient injury or surgical delay has been reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MAYFIELD COMPOSITE SERIES SKULL CLAMP
Type of Device
¿COMPOSITE SERIES¿
Manufacturer (Section D)
INTEGRA LIFESCIENCES CORPORATION OH/USA
4900 charlemar drive
4900 charlemar drive
cincinnati OH
Manufacturer (Section G)
INTEGRA LIFESCIENCES CORPORATION OH/USA
4900 charlemar drive
cincinnati OH
Manufacturer Contact
vivian nelson
1100 campus drive
princeton, NJ 
6099362319
MDR Report Key13236458
MDR Text Key290469351
Report Number3004608878-2022-00007
Device Sequence Number1
Product Code HBL
UDI-Device Identifier10381780253792
UDI-Public10381780253792
Combination Product (y/n)N
PMA/PMN Number
K142238
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type User Facility,Company Representative
Reporter Occupation Other
Type of Report Initial
Report Date 01/12/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received01/12/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberA3059
Device Catalogue NumberA3059
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/28/2021
Date Manufacturer Received12/21/2021
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured08/23/2016
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-