Model Number A1059
Device Problem Mechanical Problem (1384)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Type  malfunction  
Event Description
A facility reported that the double pin side locking mechanism of the mayfield modified skull clamp (a1059) would not lock.It is unknown if there was patient involvement; however, no patient injury or surgical delay has been reported.
 
Manufacturer Narrative
The mayfield modified skull clamp (a1059) was returned for evaluation: device history record (dhr): the dhr was reviewed and shows no abnormalities related to the reported failure.Failure analysis: the investigation of the returned device showed that the reported complaint was confirmed from the evaluation.The unit was received with the lock having rotational and lateral movement and a residue buildup present.New components were added to replace worn internal parts.Root cause analysis: complaint was confirmed via inspection of the unit.Unit received with the lock having rotational and lateral movement and required replacement of worn internal parts.No further investigation required based on the acceptability of risk and no adverse trends identified.This will be monitored and trended going forward.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MAYFIELD MODIFIED SKULL CLAMP
Type of Device
SKULL CLAMPS AND HEADREST SYSTEMS
Manufacturer (Section D)
INTEGRA LIFESCIENCES CORPORATION OH/USA
4900 charlemar drive
4900 charlemar drive
cincinnati OH
Manufacturer (Section G)
INTEGRA LIFESCIENCES CORPORATION OH/USA
4900 charlemar drive
cincinnati OH
Manufacturer Contact
vivian nelson
1100 campus drive
princeton, NJ 
6099362319
MDR Report Key13236461
MDR Text Key284529062
Report Number3004608878-2022-00005
Device Sequence Number1
Product Code HBL
UDI-Device Identifier10381780253457
UDI-Public10381780253457
Combination Product (y/n)N
PMA/PMN Number
PRE-AMEND
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type User Facility
Reporter Occupation Other
Type of Report Initial
Report Date 01/12/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received01/12/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberA1059
Device Catalogue NumberA1059
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/13/2021
Date Manufacturer Received12/20/2021
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured02/19/2021
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown