The tip of the bronchoscope tearing off was due to the user bending the et tube, causing this section of the ascope to not be in a straight position during removal of the scope.Thus, user applied excessive force to remove the scope from the tube causing the ascope bending section to tear off during removal of the scope.Further, according to the ifu for ascope 4 broncho family the user should always make sure that the bending section is in a straight position (neutral/non-deflected) and excessive force should not be used when withdrawing the endoscope.The product risk evaluation has been evaluated and the risk is assessed as acceptable.No remedial, corrective or preventive actions are taken as a result of this event.Ambu will keep monitoring this issue and take actions if necessary.
|