• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: HILL-ROM BATESVILLE TOTALCARE CONNECT SPORT+ W/MCM; BED, AC-POWERED ADJUSTABLE HOSPITAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

HILL-ROM BATESVILLE TOTALCARE CONNECT SPORT+ W/MCM; BED, AC-POWERED ADJUSTABLE HOSPITAL Back to Search Results
Model Number P1900NRENT01
Device Problem Smoking (1585)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 12/20/2021
Event Type  malfunction  
Manufacturer Narrative
The hillrom technician found the damaged power cord needed to be replaced.Per the hillrom service manual the totalcare bed system requires an effective maintenance program.We recommend that you perform a semi-annual preventative maintenance.Preventative maintenance will minimize downtime due to excessive wear.Examine the power cords and plugs for cuts, nicks, breaks, frays and for correct grounding.A search of the hillrom maintenance records showed hillrom performed preventative maintenance on this bed in (b)(6) 2021.It is unknown if the facility performed any other preventative maintenance on this bed.The technician replaced the damaged power cord to resolve the issue.Based on this information, no further action is required.
 
Event Description
Hillrom received a report from a hillrom technician stating the bed power cord had smoking and smelled electric burning.The bed was located at the account.There was no patient/user injury reported.This report was filed in our complaint handling system as complaint #(b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TOTALCARE CONNECT SPORT+ W/MCM
Type of Device
BED, AC-POWERED ADJUSTABLE HOSPITAL
Manufacturer (Section D)
HILL-ROM BATESVILLE
1069 state route 46 east
batesville IN 47006
Manufacturer Contact
harish vishwanathan
1069 state route 46 east
batesville, IN 47006
3127289851
MDR Report Key13255533
MDR Text Key283808069
Report Number1824206-2022-00026
Device Sequence Number1
Product Code FNL
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K962942
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type User Facility,Company Representative
Reporter Occupation Other
Type of Report Initial
Report Date 01/13/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received01/13/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Model NumberP1900NRENT01
Was Device Available for Evaluation? No
Was the Report Sent to FDA? No
Date Manufacturer Received12/20/2021
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured11/05/2014
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-