MAUDE Adverse Event Report: MAKO SURGICAL CORP. MAKO X3 UNI ONLAY TIBIAL INSERT SIZE 2 - 8 MM; PROSTHESIS, KNEE PATELLOFEMOROTIBIAL, PARTIAL, SEMI-CONSTRAINED, CEMENTED, POLYM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MAKO SURGICAL CORP. MAKO X3 UNI ONLAY TIBIAL INSERT SIZE 2 - 8 MM; PROSTHESIS, KNEE PATELLOFEMOROTIBIAL, PARTIAL, SEMI-CONSTRAINED, CEMENTED, POLYM Back to Search Results
Model Number 180732-1
Device Problem Difficult to Insert (1316)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 12/23/2021
Event Type  malfunction  
Event Description
Poly has an unnatural lip on it, was not able to properly seat in it.No surgical delay.Case type / application: pka (mics).Update: another insert was able to be inserted into the baseplate and implanted into the patient without issue.
 
Manufacturer Narrative
An event regarding seating/ locking issues involving a mako insert was reported.The event was not confirmed.Method & results: product evaluation and results: not performed as the device was not returned.Clinician review: no medical records were received for review with a clinical consultant.Product history review: review of the device history records indicate devices were manufactured and accepted into final stock with no relevant reported discrepancies.Complaint history review: there have been no other similar events for the lot referenced.Conclusion: the exact cause of the event could not be determined because insufficient information was provided.Further information such as photographs of the event and return of the device are needed to complete the investigation for determining root cause.No further investigation for this event is possible at this time.If devices and / or additional information become available to indicate further evaluation is warranted, this record will be reopened.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MAKO X3 UNI ONLAY TIBIAL INSERT SIZE 2 - 8 MM
Type of Device
PROSTHESIS, KNEE PATELLOFEMOROTIBIAL, PARTIAL, SEMI-CONSTRAINED, CEMENTED, POLYM
Manufacturer (Section D)
MAKO SURGICAL CORP.
2555 davie road
fort lauderdale FL 33317
Manufacturer (Section G)
STRYKER ORTHOPAEDICS-MAHWAH
325 corporate drive
mahwah NJ 07430
Manufacturer Contact
brad curtis
325 corporate drive
mahwah, NJ 07430
2018315000
MDR Report Key13257062
MDR Text Key283837110
Report Number3005985723-2022-00007
Device Sequence Number1
Product Code NPJ
UDI-Device Identifier00848486016395
UDI-Public00848486016395
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K150307
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Company Representative
Reporter Occupation Other
Type of Report Initial
Report Date 01/14/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received01/14/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number180732-1
Device Catalogue Number180732-1
Device Lot NumberX06099
Was Device Available for Evaluation? No
Date Manufacturer Received12/23/2021
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured09/16/2020
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Other;
-
-