MAKO SURGICAL CORP. MAKO X3 UNI ONLAY TIBIAL INSERT SIZE 2 - 8 MM; PROSTHESIS, KNEE PATELLOFEMOROTIBIAL, PARTIAL, SEMI-CONSTRAINED, CEMENTED, POLYM
|
Back to Search Results |
|
Model Number 180732-1 |
Device Problem
Difficult to Insert (1316)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 12/23/2021 |
Event Type
malfunction
|
Event Description
|
Poly has an unnatural lip on it, was not able to properly seat in it.No surgical delay.Case type / application: pka (mics).Update: another insert was able to be inserted into the baseplate and implanted into the patient without issue.
|
|
Manufacturer Narrative
|
An event regarding seating/ locking issues involving a mako insert was reported.The event was not confirmed.Method & results: product evaluation and results: not performed as the device was not returned.Clinician review: no medical records were received for review with a clinical consultant.Product history review: review of the device history records indicate devices were manufactured and accepted into final stock with no relevant reported discrepancies.Complaint history review: there have been no other similar events for the lot referenced.Conclusion: the exact cause of the event could not be determined because insufficient information was provided.Further information such as photographs of the event and return of the device are needed to complete the investigation for determining root cause.No further investigation for this event is possible at this time.If devices and / or additional information become available to indicate further evaluation is warranted, this record will be reopened.
|
|