Bedside nurse was bathing the patient, and noticed blood on the sheets of the bed.No alarms were active, and the pump was performing as usual.Upon further inspection, blood was coming from the outflow portion of the device.The nurse notified the charge nurse, physician, and circulatory support.The tubing was put secured by circulatory support and reinforced with a zip tie band on the outflow portion of the device.Ongoing monitoring of patient continued with no further pump concerns.Manufacturer's clinical representative was called, and surveyed the incident.Approved the securing of the outflow portion with a tie band.The observed portion of tubing that was noted to be leaking is preinstalled by the manufacturer.This is the amended submission for the previous form (b)(4).Fda safety report id# (b)(4).
|