Model Number 978A128 |
Device Problems
Human-Device Interface Problem (2949); Insufficient Information (3190)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 01/10/2022 |
Event Type
malfunction
|
Event Description
|
Information was received from healthcare provider via representative regarding a patient who was implanted with an implantable neurostimulator (ins) for unknown indications for use.It was reported that when doctor pulled lead out of pocket to disconnect from the extension , the lead just slid out of the extension hub without use of the torque wrench.The impedance check in pre-op yielded all green impedances and patient felt stim in anus and low buttocks at less than 1.0ma throughout the trial on several different programs; stage 2 lead pull today (b)(6); stage 1 lead implant was (b)(6).Trial was from (b)(6) to (b)(6); usage graph on ens showed patient had therapy on 24 hours/day.There was no surgical intervention that occurred.The issue was resolved at the time of the report.
|
|
Manufacturer Narrative
|
Concomitant medical products: product id: 3560030, serial#: (b)(4), product type: extension.Other relevant device(s) are: product id: 3560030, serial/lot #: (b)(4), ubd: 12-jul-2023, udi#: (b)(4).If information is provided in the future, a supplemental report will be issued.
|
|
Manufacturer Narrative
|
H3: the returned lead and extension were subjected to a series of standard tests that include but is not limited to visual inspection and electrical testing.The returned devices passed all testing in the laboratory and no anomalies were identified.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
|
Search Alerts/Recalls
|
|
|