MAUDE Adverse Event Report: TOSOH HI-TEC, INC. AIA-360; FLUOROMETER, FOR CLINICAL USE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

TOSOH HI-TEC, INC. AIA-360; FLUOROMETER, FOR CLINICAL USE Back to Search Results
Model Number AIA-360 IMMUNOASSAY ANALYZER
Device Problems Infusion or Flow Problem (2964); Material Twisted/Bent (2981)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 12/29/2021
Event Type  malfunction  
Manufacturer Narrative
A field service engineering (fse) was at the customer's site to address reported event.Fse confirmed the complaint by reviewing the error logs and reproduced error by priming substrate.While troubleshooting, fse noticed during the first prime for substrate, the wash portion of the prime did not dispense any wash.Fse inspected the wash tubing and found the wash tubing was pinched near the wash probe and adjusted the tubing.Fse observed operation during reagent prime and priming was normal without errors.Fse successfully performed daily run and quality control run without error and within acceptable range.No further action required by field service.The aia-360 analyzer is functioning as expected.(b)(4).The aia-360 training manual under tab 8: troubleshooting states the following: [2017] substrate purge failure is generated when no substrate was dispensed at daily maintenance.The operator is instructed to check substrate level and replace as necessary, repeat daily maintenance.The most probable cause of the reported event is due to pinched wash probe tubing.
 
Event Description
A customer reported getting error message "2017 substrate purge failure" on the aia-360 analyzer.The analyzer is down.A field service engineer (fse) was dispatched to address the reported event, which resulted in a delayed reporting of patient samples for beta human chorionic gonadotropin (bhcg), estradiol (e2) and progesterone (prog ii).There is no indication of any patient intervention or adverse health consequences due to the delay in reporting of patient results.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AIA-360
Type of Device
FLUOROMETER, FOR CLINICAL USE
Manufacturer (Section D)
TOSOH HI-TEC, INC.
1-37, fukugawa minami-machi
shunan, yamaguchi 74600 42
JA  7460042
Manufacturer (Section G)
TOSOH HI-TEC, INC.
1-37, fukugawa minami-machi
shunan, yamaguchi 74600 42
JA   7460042
Manufacturer Contact
bernadette o connell
shiba-koen first building
3-8-2 shiba
minato-ku 10586-23
JA   1058623
6506368143
MDR Report Key13262981
MDR Text Key290529956
Report Number3004529019-2022-00009
Device Sequence Number1
Product Code KHO
UDI-Device Identifier04560189282896
UDI-Public04560189282896
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K971103
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 01/17/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received01/16/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberAIA-360 IMMUNOASSAY ANALYZER
Device Catalogue Number019945
Device Lot NumberN/A
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Date Manufacturer Received12/29/2021
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured03/01/2019
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-