Model Number AIA-360 IMMUNOASSAY ANALYZER |
Device Problems
Infusion or Flow Problem (2964); Material Twisted/Bent (2981)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 12/29/2021 |
Event Type
malfunction
|
Manufacturer Narrative
|
A field service engineering (fse) was at the customer's site to address reported event.Fse confirmed the complaint by reviewing the error logs and reproduced error by priming substrate.While troubleshooting, fse noticed during the first prime for substrate, the wash portion of the prime did not dispense any wash.Fse inspected the wash tubing and found the wash tubing was pinched near the wash probe and adjusted the tubing.Fse observed operation during reagent prime and priming was normal without errors.Fse successfully performed daily run and quality control run without error and within acceptable range.No further action required by field service.The aia-360 analyzer is functioning as expected.(b)(4).The aia-360 training manual under tab 8: troubleshooting states the following: [2017] substrate purge failure is generated when no substrate was dispensed at daily maintenance.The operator is instructed to check substrate level and replace as necessary, repeat daily maintenance.The most probable cause of the reported event is due to pinched wash probe tubing.
|
|
Event Description
|
A customer reported getting error message "2017 substrate purge failure" on the aia-360 analyzer.The analyzer is down.A field service engineer (fse) was dispatched to address the reported event, which resulted in a delayed reporting of patient samples for beta human chorionic gonadotropin (bhcg), estradiol (e2) and progesterone (prog ii).There is no indication of any patient intervention or adverse health consequences due to the delay in reporting of patient results.
|
|
Search Alerts/Recalls
|
|
|