H4: the lot was manufactured from september 07, 2021 - september 08, 2021.H10: the device was received for evaluation containing 80ml of fluid in the bladder.Visual inspection did not identify any abnormalities that could have contributed to the reported condition.After the luer cap was removed and the slide clamp was disengaged from the tubing, evidence of continuous flow was visually observed at the distal luer.A functional flow rate test was performed and the flow rate was found to be within the product specification range.The reported condition was not verified.A batch review was conducted and there were no deviations found related to this reported condition during the manufacture of this lot.Should additional relevant information become available, a supplemental report will be submitted.
|