• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ZIMMER SURGICAL, INC. ZIMMER AIR DERMATOME

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ZIMMER SURGICAL, INC. ZIMMER AIR DERMATOME Back to Search Results
Model Number N/A
Device Problems Mechanical Problem (1384); Component Missing (2306); Failure to Calibrate (2440); Material Deformation (2976)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 01/12/2022
Event Type  malfunction  
Event Description
It was reported that unit was returned for maintenance but repair was needed for damaged 2 inch and 4 inch width plates, as well as for a control bar issue, calibration issue, worn screws, worn needle bearing, damaged external retaining ring, missing hinge need to be replaced, defective swivel need to be replaced.No adverse events were reported as a result of this malfunction.
 
Manufacturer Narrative
This event has been recorded by zimmer biomet under (b)(4).Telephone number: (b)(6).Once the investigation is complete, a follow up/final report will be submitted.
 
Event Description
No additional event information available.
 
Manufacturer Narrative
This event has been recorded by zimmer biomet under (b)(4).Review of the most recent repair record determined that the control bar position was not correct, the screws and bearings were worn, the e-ring was deformed, the hinge gasket was missing, and the swivel was loose.The unit was recalibrated, and the screws, bearings, e-ring, the hinge gasket, and swivel were replaced to resolve the reported issue.The width plates were damaged, however, they were returned back to the account without repairs, as the component can only be replaced.Review of the device history record identified no deviations or anomalies during manufacturing related to the reported event.A definitive root cause cannot be determined.The event is confirmed.If any further information is found which would change or alter any conclusions or information, a supplemental report will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ZIMMER AIR DERMATOME
Type of Device
DERMATOME
Manufacturer (Section D)
ZIMMER SURGICAL, INC.
200 west ohio avenue
dover OH 44622
Manufacturer (Section G)
ZIMMER SURGICAL, INC.
200 west ohio avenue
dover OH 44622
Manufacturer Contact
christina arnt
56 e. bell drive
warsaw, IN 46582
5745273773
MDR Report Key13308020
MDR Text Key284388166
Report Number0001526350-2022-00050
Device Sequence Number1
Product Code GFD
Combination Product (y/n)N
Reporter Country CodeUK
PMA/PMN Number
EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,User Facility
Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date 04/14/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received01/20/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberN/A
Device Catalogue Number00880100100
Device Lot Number60720310
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/12/2022
Was the Report Sent to FDA? No
Date Manufacturer Received03/29/2022
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured06/30/2007
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Removal/Correction NumberN/A
Patient Sequence Number1
-
-