Model Number N/A |
Device Problems
Mechanical Problem (1384); Component Missing (2306); Failure to Calibrate (2440); Material Deformation (2976)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 01/12/2022 |
Event Type
malfunction
|
Event Description
|
It was reported that unit was returned for maintenance but repair was needed for damaged 2 inch and 4 inch width plates, as well as for a control bar issue, calibration issue, worn screws, worn needle bearing, damaged external retaining ring, missing hinge need to be replaced, defective swivel need to be replaced.No adverse events were reported as a result of this malfunction.
|
|
Manufacturer Narrative
|
This event has been recorded by zimmer biomet under (b)(4).Telephone number: (b)(6).Once the investigation is complete, a follow up/final report will be submitted.
|
|
Event Description
|
No additional event information available.
|
|
Manufacturer Narrative
|
This event has been recorded by zimmer biomet under (b)(4).Review of the most recent repair record determined that the control bar position was not correct, the screws and bearings were worn, the e-ring was deformed, the hinge gasket was missing, and the swivel was loose.The unit was recalibrated, and the screws, bearings, e-ring, the hinge gasket, and swivel were replaced to resolve the reported issue.The width plates were damaged, however, they were returned back to the account without repairs, as the component can only be replaced.Review of the device history record identified no deviations or anomalies during manufacturing related to the reported event.A definitive root cause cannot be determined.The event is confirmed.If any further information is found which would change or alter any conclusions or information, a supplemental report will be filed accordingly.Zimmer biomet will continue to monitor for trends.
|
|
Search Alerts/Recalls
|
|
|