Catalog Number CAR-02800
Device Problem Crack (1135)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 10/22/2021
Event Type  malfunction  
Event Description
It was reported the luer hub was found cracked during hemodialysis.No patient harm reported.The patient's condition is reported as fine.
 
Manufacturer Narrative
(b)(4).
 
Manufacturer Narrative
Qn#: (b)(4).Complaint verification testing could not be performed as no sample was returned for analysis.A device history record review was performed and no relevant findings were identified.Without the device to evaluate the complaint could not be confirmed and the probable cause could not be determined from the available information.Teleflex will continue to monitor and trend for reports of this nature.If the sample becomes available at a later date a follow up report will be submitted with investigation results.
 
Event Description
It was reported the luer hub was found cracked during hemodialysis.No patient harm reported.The patient's condition is reported as fine.
 
Event Description
It was reported the luer hub was found cracked during hemodialysis.No patient harm reported.The patient's condition is reported as fine.
 
Manufacturer Narrative
(b)(4).The customer returned a connector assembly for evaluation.Signs of use in the form of biological material were observed inside both extension lines.Visual analysis revealed the venous (blue) luer hub was cracked and contained scratches on the body.The damage appeared consistent with repeated tightening on the luer.The instructions for use (ifu) provided with this kit states, "flush both lumens of catheter with saline and clamp irrigation tube with catheter pinch clamp".The connector assembly was functionally tested by injecting both lumens using a water-filled lab inventory 10ml syringe.Water leaked from the cracked hub when injected with water.The instructions for use (ifu) provided with this kit warns the user, "repeated over tightening of bloodlines, syringes and caps will reduce connector life and could lead to potential connector failure".The customer report of a cracked luer hub was confirmed by complaint investigation of the returned sample.Visual analysis revealed that the venous luer hub was cracked and contained multiple scratches.The appearance of the crack was consistent with over-tightening on the luer hub.A device history record review was performed with no relevant findings.Based on the customer report and the condition of the sample received, unintentional use error caused or contributed to this event.Teleflex will continue to monitor and trend for complaints of this nature.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ARROW CANNON II PLUS REPLACEMENT HUB SET
Type of Device
KIT, REPAIR, CATHETER, HEMODI
Manufacturer (Section D)
ARROW INTERNATIONAL INC.
reading PA
Manufacturer (Section G)
ARROW INTERNATIONAL DE MEXICO S.A. DE C.V.
ave. washington 3701
colonia panamericana, chihuahua
chihuahua 31200
MX   31200
Manufacturer Contact
katharine tarpley
3015 carrington mill blvd
morrisville, NC 27560
MDR Report Key13310220
MDR Text Key286594170
Report Number9680794-2022-00008
Device Sequence Number1
Product Code NFK
Combination Product (y/n)N
Reporter Country CodeCH
PMA/PMN Number
K020430
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,User Facility,Company Representative,Distributor
Reporter Occupation Other
Type of Report Initial,Followup,Followup
Report Date 12/29/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received01/20/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Expiration Date03/31/2023
Device Catalogue NumberCAR-02800
Device Lot Number13F20D0293
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? No
Date Manufacturer Received04/27/2022
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured03/31/2023
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
HEMODIALYSIS.; HEMODIALYSIS.; HEMODIALYSIS.
-
-