Model Number 989605463521 |
Device Problem
Incorrect Measurement (1383)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 12/22/2021 |
Event Type
malfunction
|
Event Description
|
A customer reported their innosight ultrasound system had produced incorrect cardiac measurements for teicholtz calculations in m mode.The issue was obvious to the user and there was no harm to the user or patient associated with this event.
|
|
Manufacturer Narrative
|
An investigation is underway to determine the root cause of the issue.Results of the investigation will be included in a follow up report upon its completion.
|
|
Manufacturer Narrative
|
According to the product defect report initiated by philips, the software involved in this issue is owned by a 3rd party supplier, and they confirmed a software bug was the root cause of the issue and has been corrected in a subsequent software version.
|
|