U.S. Department of Health and Human Services

MAUDE Adverse Event Report: PHILIPS ULTRASOUND INNOSIGHT; SYSTEM, IMAGING, PULSED ECHO, ULTRASONIC

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

PHILIPS ULTRASOUND INNOSIGHT; SYSTEM, IMAGING, PULSED ECHO, ULTRASONIC Back to Search Results
Model Number 989605463521
Device Problem Incorrect Measurement (1383)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 12/22/2021
Event Type  malfunction  
Event Description
A customer reported their innosight ultrasound system had produced incorrect cardiac measurements for teicholtz calculations in m mode.The issue was obvious to the user and there was no harm to the user or patient associated with this event.
 
Manufacturer Narrative
An investigation is underway to determine the root cause of the issue.Results of the investigation will be included in a follow up report upon its completion.
 
Manufacturer Narrative
According to the product defect report initiated by philips, the software involved in this issue is owned by a 3rd party supplier, and they confirmed a software bug was the root cause of the issue and has been corrected in a subsequent software version.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INNOSIGHT
Type of Device
SYSTEM, IMAGING, PULSED ECHO, ULTRASONIC
Manufacturer (Section D)
PHILIPS ULTRASOUND
22100 bothell everett highway
bothell WA 98021
Manufacturer (Section G)
PHILIPS ULTRASOUND
22100 bothell everett highway
bothell WA 98021
Manufacturer Contact
eric nail
22100 bothell everett highway
bothell, WA 98021
MDR Report Key13311285
MDR Text Key286951199
Report Number3019216-2022-00008
Device Sequence Number1
Product Code IYN
Combination Product (y/n)N
Reporter Country CodeSP
PMA/PMN Number
K132304
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received01/20/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number989605463521
Device Catalogue Number795001
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Was Device Evaluated by Manufacturer? No
Date Device Manufactured03/15/2021
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-