This report is being supplemented to provide additional information based on the legal manufacturer's final investigation.Please see updates in a1, d4, d8, e2, g2, h4, and h6.Additionally, g3 - aware date in original mdr should be (b)(6) 2021.A review of the device history record found no deviations that could have caused or contributed to the reported issue.Based on the results of the investigation, it¿s likely that either water supply valve did not work due to a temporary poor continuity and the water supply could not stop, or the water supply valve could not be controlled and the water supply could not be stopped causing damage to the electrical circuit board.The specific root cause of the event was unable to be identified.Olympus will continue to monitor field performance for this device.
|