Model Number MR290V
Device Problems Corroded (1131); Fluid/Blood Leak (1250); Leak/Splash (1354)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 01/05/2022
Event Type  malfunction  
Event Description
A distributor reported on behalf of a healthcare facility in (b)(6) via a fisher and paykel healthcare (f&p) field representative, that a mr290v vented autofeed humidification chamber was leaking water during patient use.There was no reported patient consequence.
 
Manufacturer Narrative
(b)(4).The complaint mr290v vented autofeed humidification chamber is currently en route to fisher & paykel healthcare (f&p) for evaluation.We will provide a follow-up report upon completion of our investigation.
 
Event Description
A distributor reported on behalf of a healthcare facility in japan via a fisher and paykel healthcare (f&p) field representative, that a mr290v vented autofeed humidification chamber was leaking water during patient use.The healthcare facility further reported that bisolvon and saline were mixed and the nebulizer was used.There was no reported patient consequence.
 
Manufacturer Narrative
(b)(4).Method: the complaint mr290v vented autofeed humidification chamber was received at fisher & paykel healthcare (f&p) in new zealand where it was visually inspected and analysed.Results: visual inspection revealed a hole was found on the base of the chamber.It was reported by the healthcare facility that bisolvon and saline were mixed and the nebulizer was used.Conclusion: the degradation of the aluminium baseplate is most likely due to the presence of saline solution in the mr290v vented autofeed humidification chamber.Saline solution contains sodium salt which is highly corrosive to aluminum.The presence of this solution could cause degradation of the aluminum chamber base over time.Every mr290v chamber is pressure tested following the manufacturing process to check for any leaks present due to cracks and other causes.Any chamber which fails this test is rejected.In addition, the pressure test is followed by a visual inspection of each chamber.The subject chamber would have met the required specification prior to distribution.Our user instructions that accompany the mr290v vented autofeed humidification chamber states the following: - "do not soak, wash, sterilize, or reuse this product.Avoid contact with chemicals, cleaning agents, or hand sanitizers."; - "use usp sterile water for inhalation or equivalent.".
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VENTED AUTOFEED HUMIDIFICATION CHAMBER
Type of Device
AUTOFEED CHAMBER
Manufacturer (Section D)
FISHER & PAYKEL HEALTHCARE LIMITED
15 maurice paykel place
east tamaki
auckland, 2013
NZ  2013
Manufacturer (Section G)
FISHER & PAYKEL HEALTHCARE LIMITED
15 maurice paykel place
east tamaki
auckland, 2013
NZ   2013
Manufacturer Contact
faranak gomarooni
173 technology dr.
suite 100
irvine, CA 92618
9494534000
MDR Report Key13330300
MDR Text Key286446547
Report Number9611451-2022-00025
Device Sequence Number1
Product Code BTT
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
K934140
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,User Facility,Company Representative,Distributor
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 01/07/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received01/23/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberMR290V
Device Catalogue NumberMR290V
Device Lot NumberNOT PROVIDED
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/24/2022
Is the Reporter a Health Professional? No
Date Manufacturer Received01/28/2022
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
F&P RT380 ADULT DUAL HEATED BREATHING CIRCUIT.; F&P RT380 ADULT DUAL HEATED BREATHING CIRCUIT.
-
-