Catalog Number 62422602
Device Problem Break (1069)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Type  malfunction  
Manufacturer Narrative
There has been a previous report received where this malfunction resulted in a serious injury.Therefore, it must be presumed that recurrence of this malfunction could possibly cause or contribute to a serious injury or require medical or surgical intervention to preclude such.As such, this event is reportable per 21cfr part 803.The device is available for evaluation, though results are not available as of this report.Evaluation results will be submitted as they become available.
 
Event Description
In this event it has been reported that automate iii broke during use.No injury to patient.
 
Manufacturer Narrative
Returned product was automate iii handle and 1 flexshaft all date coded 1220 (manufactured dec 2020) with weld failure and/or coil deformation causing the product to not function properly.Capa-2021-375 opened to address weld failures for product manufactured by sarasota since december 2020 (date code 1220).Complaint is considered substantiated.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AUTOMATE III
Type of Device
INSTRUMENTS, DENTAL HAND
Manufacturer (Section D)
DENTSPLY LLC
38 west clarke avenue
milford DE 19963
Manufacturer (Section G)
DENTSPLY LLC
38 west clarke avenue
milford DE 19963
Manufacturer Contact
hannah seevaratnam
221 west philadelphia st.
york, PA 17401
7178457511
MDR Report Key13333888
MDR Text Key286596851
Report Number2515379-2022-00032
Device Sequence Number1
Product Code DZN
Combination Product (y/n)N
Reporter Country CodeIT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional
Reporter Occupation Dentist
Type of Report Initial,Followup
Report Date 06/08/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received01/24/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Catalogue Number62422602
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/18/2022
Is the Reporter a Health Professional? Yes
Distributor Facility Aware Date01/13/2022
Date Manufacturer Received01/13/2022
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Type of Device Usage A
Patient Sequence Number1
-
-