• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BAXTER HEALTHCARE CORPORATION CARTRIDGE; SET, TUBING, BLOOD, WITH AND WITHOUT ANTI-REGURGITATION VALVE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BAXTER HEALTHCARE CORPORATION CARTRIDGE; SET, TUBING, BLOOD, WITH AND WITHOUT ANTI-REGURGITATION VALVE Back to Search Results
Catalog Number 101025
Device Problems Fluid/Blood Leak (1250); Loose or Intermittent Connection (1371)
Patient Problem Appropriate Clinical Signs, Symptoms, Conditions Term / Code Not Available (4581)
Event Date 12/27/2021
Event Type  malfunction  
Event Description
It was reported that within seven minutes of starting treatment with a gambro cartridge, the venous line started leaking out blood.It was further described that the venous leur lock, from the blood cartridge, appeared to be tightened, however, when rotated further, it became loose.The amount of blood loss was reported to be approximately 5ml.A prophylactic treatment with vancomycin was administered intravenously.There was no patient injury or medical intervention associated with this event.No additional information is available.
 
Manufacturer Narrative
Should additional relevant information become available, a supplemental report will be submitted.
 
Manufacturer Narrative
H10: the actual device was not available; however, one (1) photograph and eight (8) retained samples were provided for evaluation.Visual inspection of the picture identified a damage in the venous patient connector.The reported condition was verified.The cause of the condition was determined to be a manufacturing issue.A nonconformance has been opened to address this issue.Visual inspection of the eight (8) retained samples, leak testing and functional tests did not identify any abnormalities that could have contributed to the reported condition.A batch review was conducted and there were no deviations found related to this reported condition during the manufacture of this lot.Should additional relevant information become available, a supplemental report will be submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CARTRIDGE
Type of Device
SET, TUBING, BLOOD, WITH AND WITHOUT ANTI-REGURGITATION VALVE
Manufacturer (Section D)
BAXTER HEALTHCARE CORPORATION
deerfield IL
Manufacturer (Section G)
BAXTER HEALTHCARE - TIJUANA BAJA
blvd pacifico 10014
parque industrial pacifico
tijuana baja california CP 22 643
MX   CP 22643
Manufacturer Contact
25212 w. illinois route 120
round lake, IL 
2242702068
MDR Report Key13335390
MDR Text Key288809599
Report Number8030638-2022-00002
Device Sequence Number1
Product Code FJK
UDI-Device Identifier07332414007836
UDI-Public(01)07332414007836
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K070414
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Company Representative
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 03/01/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received01/24/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number101025
Device Lot NumberMA0121166001
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Date Manufacturer Received03/01/2022
Was Device Evaluated by Manufacturer? No
Date Device Manufactured06/29/2021
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-