NOVA BIOMEDICAL CORP. NOVA STATSTRIP HOSPITAL METER SYSTEM; GLUCOSE TEST SYSTEM Back to Search Results
Model Number 1.75
Device Problem Patient Data Problem (3197)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 12/28/2021
Event Type  malfunction  
Manufacturer Narrative
There is currently a pending investigation.Nova is requesting additional information and further details will be provided in a supplemental report.
 
Event Description
The customer reports that a statstrip meter scanned a patient armband incorrectly.When one patient was scanned a different patient's data appeared.After a rescan the correct patient's information appeared so the meter's serial number was not taken note of.There was no patient harm or intervention reported.
 
Manufacturer Narrative
Nova initiated an investigation into this meter mis-scan.By requesting, the patient armbands in question.The customer sent sample armbands.And they were graded using a barcode verifier.The quality of the armbands was very low.And it was concluded, that the mis-scan was caused by the poor quality of the armbands.The customer is reviewing their printers and settings, but the nova meter performed as it should.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
NOVA STATSTRIP HOSPITAL METER SYSTEM
Type of Device
GLUCOSE TEST SYSTEM
Manufacturer (Section D)
NOVA BIOMEDICAL CORP.
200 prospect street
waltham MA 02453 9141
Manufacturer (Section G)
NOVA BIOMEDICAL CORP.
200 prospect street
waltham MA 02453 9141
Manufacturer Contact
katrina thorne gates
200 prospect street
waltham, MA 02453-9141
7816473700
MDR Report Key13359388
MDR Text Key287129732
Report Number1219029-2022-00006
Device Sequence Number1
Product Code PZI
UDI-Device Identifier00385480533980
UDI-Public00385480533980
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K181043
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,User Facility
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 03/09/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received01/26/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number1.75
Device Catalogue Number53398
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received02/11/2022
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-