Catalog Number 110003452
Device Problem Difficult or Delayed Separation (4044)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 01/04/2022
Event Type  malfunction  
Manufacturer Narrative
(b)(4).Product has been received by zimmer biomet and the investigation is in process.Once the investigation has been completed, a follow-up mdr will be submitted.Concomitant medical products: item # 010002736/ g7 ball hex drvr for insr hndl/ lot # 062100; item # 110010247/ g7 osseoti 4 hole shell 58mm/ lot # unknown; item # 110003451/ g7 str modular shell inserter/ lot # 494810.Multiple mdr reports were filed for this event, please see associated reports: 0001825034 -2022 -00195, 0001825034 -2022 -00197 and 0001825034 -2022 -00199.
 
Event Description
It was reported that during a hip procedure the cup insertion handle could not be loosened from cup.Tommy bar was used with g7 ball tip screw driver to try to release cup.G7 ball driver broke at tip and insertion handle was still attached to cup in patient.Whole construct was removed.No adverse events were reported.Attempts have been made and additional information on the reported event is unavailable at this time.
 
Event Description
No additional event information to report at this time.
 
Manufacturer Narrative
This follow-up report is being submitted to relay additional information.Visual inspection of the returned threaded shaft found wear rings and scratching on the shaft.The large hex feature is dinged and scratched.The lead thread on the tip has been deformed.Dhr was reviewed and no discrepancies were found.The root cause is unable to be determined.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
G7 STR INSRTR THREADED SHAFT
Type of Device
INSTRUMENT, HIP
Manufacturer (Section D)
ZIMMER BIOMET, INC.
56 e. bell drive
warsaw IN 46582
Manufacturer (Section G)
ZIMMER BIOMET, INC.
56 e. bell drive
warsaw IN 46582
Manufacturer Contact
christina arnt
56 e. bell dr.
warsaw, IN 46582
5745273773
MDR Report Key13368079
MDR Text Key286362164
Report Number0001825034-2022-00200
Device Sequence Number1
Product Code PBI
UDI-Device Identifier00880304551602
UDI-Public(01)00880304551602(11)201105(10)072082
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K121874
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,Distributor
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 03/22/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received01/27/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number110003452
Device Lot Number072082
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/19/2022
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received03/18/2022
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured11/05/2020
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Removal/Correction NumberN/A
Patient Sequence Number1
Patient Age61 YR
Patient SexMale
-
-