TERUMO CARDIOVASCULAR SYSTEMS CORPORATION ADVANCED PERFUSION SYSTEM 1; CONSOLE, HEART-LUNG MACHINE, CARDIOPULMONARY BYPASS- ADVANCED PERFUSION SYSTEM 1
|
Back to Search Results |
|
Model Number 801188 |
Device Problem
Fluid/Blood Leak (1250)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 01/05/2022 |
Event Type
malfunction
|
Manufacturer Narrative
|
The reported complaint was confirmed.The srt observed above 10 cubic centimeters per minute (cc/min) there was a leak during the epgs leak test.The brass plumbing fitting was found to be leaking when the liquid leak detector was applied.The fitting was replaced.The unit operated to the manufacturer's specifications.If additional information becomes available on this complaint that would alter the facts and/or conclusion, a supplemental report will be filed accordingly.
|
|
Event Description
|
The service repair technician (srt) reported that during testing of the device at the service center, the brass plumbing fitting on the electronic patient gas system (epgs) was leaking.There was no patient involvement.
|
|