MAUDE Adverse Event Report: BAXTER HEALTHCARE CORPORATION HOMECHOICE PRO; SYSTEM, PERITONEAL, AUTOMATIC DELIVERY

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BAXTER HEALTHCARE CORPORATION HOMECHOICE PRO; SYSTEM, PERITONEAL, AUTOMATIC DELIVERY Back to Search Results
Catalog Number 5C8310
Device Problem Filling Problem (1233)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 01/03/2022
Event Type  malfunction  
Event Description
It was reported that a homechoice pro underdelivered peritoneal dialysis (pd) solution to the patient.The caregiver (cg) reported that the expected delivery volume is 2500ml; however, patient was filled with 2200ml.This issue was further described as, ¿the cycler is not filling as per the prescription and there is too much solution left over at the end of therapy".This was observed during use of the device for peritoneal dialysis therapy.Renal therapy services (rts) assisted the caregiver (cg) to reset the procard and verified the prescription, check the therapy settings, and explained the issue.There was no report of patient injury or medical intervention associated with this event.No additional information is available.
 
Manufacturer Narrative
Should additional relevant information become available, a supplemental report will be submitted.
 
Manufacturer Narrative
The device was received for evaluation.The event history log review showed no keystrokes, programming, or use related events that indicated and/or contributed to the reported issue.Visual inspection did not identify any abnormalities that could have contributed to the reported condition.The returned device passed functional testing and was found to meet specification.A service history review revealed no indication that the parts replaced during servicing caused or contributed to the reported event.The reported condition was not verified.Should additional relevant information become available, a supplemental report will be submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HOMECHOICE PRO
Type of Device
SYSTEM, PERITONEAL, AUTOMATIC DELIVERY
Manufacturer (Section D)
BAXTER HEALTHCARE CORPORATION
deerfield IL
Manufacturer (Section G)
BAXTER HEALTHCARE - SINGAPORE
2 woodlands industrial park d
singapore 73875 0
SN   738750
Manufacturer Contact
25212 w. illinois route 120
round lake, IL 
2242702068
MDR Report Key13388750
MDR Text Key284677782
Report Number1416980-2022-00198
Device Sequence Number1
Product Code FKX
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K102936
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Consumer,Company Representative
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 02/25/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received01/28/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Catalogue Number5C8310
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/12/2022
Is the Reporter a Health Professional? No
Date Manufacturer Received02/23/2022
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
UNKNOWN PD SOLUTION.
-
-