Catalog Number 5C8310 |
Device Problem
Filling Problem (1233)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 01/03/2022 |
Event Type
malfunction
|
Event Description
|
It was reported that a homechoice pro underdelivered peritoneal dialysis (pd) solution to the patient.The caregiver (cg) reported that the expected delivery volume is 2500ml; however, patient was filled with 2200ml.This issue was further described as, ¿the cycler is not filling as per the prescription and there is too much solution left over at the end of therapy".This was observed during use of the device for peritoneal dialysis therapy.Renal therapy services (rts) assisted the caregiver (cg) to reset the procard and verified the prescription, check the therapy settings, and explained the issue.There was no report of patient injury or medical intervention associated with this event.No additional information is available.
|
|
Manufacturer Narrative
|
Should additional relevant information become available, a supplemental report will be submitted.
|
|
Manufacturer Narrative
|
The device was received for evaluation.The event history log review showed no keystrokes, programming, or use related events that indicated and/or contributed to the reported issue.Visual inspection did not identify any abnormalities that could have contributed to the reported condition.The returned device passed functional testing and was found to meet specification.A service history review revealed no indication that the parts replaced during servicing caused or contributed to the reported event.The reported condition was not verified.Should additional relevant information become available, a supplemental report will be submitted.
|
|