MAQUET CARDIOPULMONARY GMBH CARDIOHELP; CONSOLE, HEART-LUNG MACHINE, CARDIOPULMONARY BYPASS Back to Search Results
Model Number CARDIOHELP-I
Device Problem Increase in Pressure (1491)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 01/27/2022
Event Type  malfunction  
Manufacturer Narrative
A follow-up emdr will be submitted when additional information becomes available.A getinge technician will investigate the cardiohelp.
 
Event Description
It was reported that the delta p reading was wrong.The cardiohelp was exchanged during treatment.No patient harm was reported.Complaint id# (b)(4).
 
Manufacturer Narrative
It was reported that the delta p reading was wrong and the device was exchanged during treatment.A getinge service technician (fst) was sent for investigation and repair on 2022-02-21/22/23.The reported failure could be confirmed.In order to rule out the connection cable for disposables (hls cable) was dirty or damaged it was replaced, but did not solve the reported issue.After replacing the sensor panel the failure was solved.The technician performed safety, calibration, and functionality checks to factory specifications.Another sensor panel with a similar failure was investigated by life cycle engineering (lce).Following could be determined: the most probable root cause is a mechanical damage of the compensated choke.Based on the results the reported failure "delta p reading wrong" could be confirmed.The occurrence rate was calculated for the reported issue and it was determined that this is not a systemic issue.Therefore, no remedial action is required.The occurrence rate related to the reported issue is currently being monitored as part of maquet cardiopulmonary¿ s trending program and additional investigations or corrections will be implemented in case of adverse trending.
 
Event Description
Complaint id# (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CARDIOHELP
Type of Device
CONSOLE, HEART-LUNG MACHINE, CARDIOPULMONARY BYPASS
Manufacturer (Section D)
MAQUET CARDIOPULMONARY GMBH
neue rottenburger strasse 37
hechingen
Manufacturer (Section G)
JULIA KAPFENBERGER
neue rottenburger strasse 37
hechingen
Manufacturer Contact
neue rottenburger strasse 37
hechingen 
MDR Report Key13391053
MDR Text Key284675155
Report Number8010762-2022-00026
Device Sequence Number1
Product Code DTQ
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K133598
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,User Facility
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 02/24/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received01/31/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberCARDIOHELP-I
Device Catalogue Number701048012
Is the Reporter a Health Professional? Yes
Date Manufacturer Received02/22/2022
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured12/01/2016
Is the Device Single Use? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-