Catalog Number 2C2117K |
Device Problem
No Flow (2991)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 01/07/2022 |
Event Type
malfunction
|
Manufacturer Narrative
|
The device has been received and the evaluation is in progress.Should additional relevant information become available, a supplemental report will be submitted.
|
|
Event Description
|
It was reported that a small volume intermate would not infuse during patient infusion.The device contained colistin 2 million iu in 54 ml sodium chloride 0.9% bp.There was no report of patient injury or medical intervention associated with this event.No additional information is available.
|
|
Manufacturer Narrative
|
The device was received for evaluation.Visual inspection did not identify any abnormalities that could have contributed to the reported condition.A functional flow rate test was performed, and the flow rate was found to be within the product specification range.The reported condition was not verified.A batch review was conducted and there were no deviations found related to this reported condition during the manufacture of this lot.Should additional relevant information become available, a supplemental report will be submitted.
|
|