As reported for this event by the customer, during a therapeutic transurethral resection of the prostate (turp) procedure, when the monopolar technology was ignited, the device sparked and there was a bit of smoke.The cord detached from the end plugging into the generator.There was no harm or adverse impact to the patient.There was no impact to the outcome of the procedure.Procedure was completed with the same device.Some other unknown device was replaced.The non-olympus generator settings were at default.The device has reused more than 10 times.
|
This report is being supplemented to provide additional information based on the legal manufacturer's final investigation.A review of the device history record found no deviations that could have caused or contributed to the reported issue.It has been over 3 years since the subject device was manufactured, therefore, it is likely that the cable was used for longer than the specified 12 months.Based on the results of the investigation, it is very likely that the age-related wear in connection with repeated extreme bending or tensile loads most likely led to the breakage of single or all the wires in the cable.This can cause a voltage spike at the damaged area when the generator is activated which may result in a spark and complete disconnection of the plug.The specific cause of the cable sparking was likely due to age-related wear in connection with improper handling.The service life of the cable is limited to 12 months.After this time, the cable should no longer be used.The following information is stated in the instructions for use regarding proper handling of the device: "additionally, the cable must be checked for damage before each use and after reprocessing.By gently pulling on the plug (max.20nm), it can be determined whether the copper strand of the cable is already damaged.If the cable does not give way but remains rigid, the cable is most likely intact.In order not to shorten the service life of the cable any further, the cable should not be wound up with a loop diameter of less than 10cm and when pulling out the cable, the plug should be pulled and not the cable.".Olympus will continue to monitor field performance for this device.
|