Testing/inspection: one unit was returned for evaluation of the failure and pictures by the customer was not provide it.Unit returned was received inside of a plastic bag which is their original package sample was submitted to visual inspection base on (b)(4) rev.100 - adult disposable anesthesia breathing circuit 2 x 72"" (183cm) tubes results: customer reported circuits from the same lot have the filter missing? reviewing the sample returned, the filter is in the bag but is not installed with the elbow.Failure description must be loose filter: see pictures attached on the investigation: conclusions: based on the inspection, it can't be confirmed failure mode as missing filter but can be confirmed as loose filter.Root cause: cannot be confirmed the failure, the most probable root cause is that loose filter occurred after the product left smiths medical facilities.Action taken: no corrective actions were taken since the investigation did not reveal that the defect was caused during the manufacturing process dhr: lot 3973607 was manufactured with (b)(4) units; lot met the requirements to release the lot with no deviations identified during their manufactured.
|