• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: HOYA CORPORATION PENTAX TOKYO OFFICE PENTAX; VIDEO DUODENO SCOPE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

HOYA CORPORATION PENTAX TOKYO OFFICE PENTAX; VIDEO DUODENO SCOPE Back to Search Results
Model Number ED-3490TK
Device Problems Partial Blockage (1065); Fracture (1260)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 01/27/2022
Event Type  malfunction  
Manufacturer Narrative
Evaluation summary: if additional information becomes available, a supplemental report will be filed with the new information.This device is classified as import for export, therefore 510k is not applicable.
 
Event Description
Service found a rest of a brush inside the op-channel.Distri submitted the scope as "op-channel clogged".This event occurred at the time of during inspection.There was no report of patient harm.
 
Manufacturer Narrative
Evaluation summary: we checked the returned unit and confirmed that the remote control buttons cut.Based on the result, we concluded that it was caused due to the excessive force applied on the remote control buttons.Correction information: g6: follow up #1 h6: coding changed based on the investigation result additional information: h4: device manufacture date.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PENTAX
Type of Device
VIDEO DUODENO SCOPE
Manufacturer (Section D)
HOYA CORPORATION PENTAX TOKYO OFFICE
tsutsujigaoka 1-1-110
akishima-shi, tokyo 196-0 012
JA  196-0012
Manufacturer (Section G)
HOYA CORPORATION PENTAX TOKYO OFFICE
tsutsujigaoka 1-1-110
akishima-shi, tokyo 196-0 012
JA   196-0012
Manufacturer Contact
william goeller
3 paragon drive
montvale, NJ 07645
8004315880
MDR Report Key13431316
MDR Text Key289504377
Report Number9610877-2022-00173
Device Sequence Number1
Product Code FDT
Combination Product (y/n)N
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Company Representative,Distributor
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 03/11/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received02/02/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberED-3490TK
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? No
Date Manufacturer Received01/27/2022
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured10/03/2013
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-