It was reported that tip detached and withdrawal difficulty occurred.The target lesion was located in the iliac vein of the lower limb.An angiojet solent omni catheter was used in a thrombectomy procedure.During the procedure under thrombectomy mode, the device alerted that the catheter needed to be re-primed.The catheter could not be withdrawn normally after trying to remove.After the catheter was simply pulled out, it was noted that the tip was detached in the sheath.The procedure was completed with another of the same device.There were no patient complications reported and patient's status was stable.
|
E1: initial reporter facility name: (b)(6) hospital.Device evaluated by mfr.: the device returned to boston scientific for analysis.The product return contained an angiojet solent omni.The assembly, supply line, shaft, saline supply and tip were all visually examined for potential damage.The catheter shaft showed a heavily stretched shaft with a detached tip.The stretching was 2.5cm long, at the initial location of the distal marker band.The proximal saddle was displaced 0.5cm from its intended location, with the marker band damaged and further displaced another 1.5cm, for a total of 2cm off the intended location.Along with this damaged marker band, a fractured hypotube was noted at the same location.A functional test was performed.The pump was placed into the ultra drive console but the device gave an under pressure error.The devices pressure was not within normal range.Due to this, a functional test cannot be performed.Testing of the device confirms the detached tip and error messages.However the difficulty in removal cannot be confirmed.
|
It was reported that tip detached and withdrawal difficulty occurred.The target lesion was located in the iliac vein of the lower limb.An angiojet solent omni catheter was used in a thrombectomy procedure.During the procedure under thrombectomy mode, the device alerted that the catheter needed to be re-primed.The catheter could not be withdrawn normally after trying to remove.After the catheter was simply pulled out, it was noted that the tip was detached in the sheath.The procedure was completed with another of the same device.There were no patient complications reported and patient's status was stable.
|