Model Number CNS-6201A |
Device Problems
Application Program Problem (2880); Power Problem (3010); Unintended Application Program Shut Down (4032)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 01/06/2022 |
Event Type
malfunction
|
Manufacturer Narrative
|
The biomedical engineer (bme) reported that the central nurse's station (cns) powered off and would not power on.They replaced the power cable and tried a wall outlet, and the cns powered back on.No patient harm was reported.Nihon kohden continues to investigate the reported event.Nihon kohden will submit a supplemental report in accordance with 21 cfr section 803.56 when additional information becomes available.Additional device information: concomitant medical device field contains no information (ni), as attempts to obtain information were made, but not provided.
|
|
Event Description
|
The biomedical engineer (bme) reported that the central nurse's station (cns) powered off and would not power on.They replaced the power cable and tried a wall outlet, and the cns powered back on.No patient harm was reported.
|
|
Event Description
|
The biomedical engineer (bme) reported that the central nurse's station (cns) powered off and would not power back on.They replaced the power cable and plugged it into a wall outlet, which resolved the issue.No patient harm was reported.
|
|
Manufacturer Narrative
|
Details of complaint: the biomedical engineer (bme) reported that the central nurse's station (cns) powered off and would not power back on.They replaced the power cable and plugged it into a wall outlet, which resolved the issue.No patient harm was reported.Investigation summary: the customer bypassed the ups and directly plugged the cns into the wall outlet, as well as replacing the power cable, and the cns turned on.Based on the available information, the possible causes of the issue are wear and tear of the ups or wear and tear of the power cable.There is no malfunction of the cns, but rather a failure of the ups or the power cable.
|
|