• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: NIHON KOHDEN CORPORATION CNS-6201A; CENTRAL MONITOR SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

NIHON KOHDEN CORPORATION CNS-6201A; CENTRAL MONITOR SYSTEM Back to Search Results
Model Number CNS-6201A
Device Problems Application Program Problem (2880); Power Problem (3010); Unintended Application Program Shut Down (4032)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 01/06/2022
Event Type  malfunction  
Manufacturer Narrative
The biomedical engineer (bme) reported that the central nurse's station (cns) powered off and would not power on.They replaced the power cable and tried a wall outlet, and the cns powered back on.No patient harm was reported.Nihon kohden continues to investigate the reported event.Nihon kohden will submit a supplemental report in accordance with 21 cfr section 803.56 when additional information becomes available.Additional device information: concomitant medical device field contains no information (ni), as attempts to obtain information were made, but not provided.
 
Event Description
The biomedical engineer (bme) reported that the central nurse's station (cns) powered off and would not power on.They replaced the power cable and tried a wall outlet, and the cns powered back on.No patient harm was reported.
 
Event Description
The biomedical engineer (bme) reported that the central nurse's station (cns) powered off and would not power back on.They replaced the power cable and plugged it into a wall outlet, which resolved the issue.No patient harm was reported.
 
Manufacturer Narrative
Details of complaint: the biomedical engineer (bme) reported that the central nurse's station (cns) powered off and would not power back on.They replaced the power cable and plugged it into a wall outlet, which resolved the issue.No patient harm was reported.Investigation summary: the customer bypassed the ups and directly plugged the cns into the wall outlet, as well as replacing the power cable, and the cns turned on.Based on the available information, the possible causes of the issue are wear and tear of the ups or wear and tear of the power cable.There is no malfunction of the cns, but rather a failure of the ups or the power cable.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CNS-6201A
Type of Device
CENTRAL MONITOR SYSTEM
Manufacturer (Section D)
NIHON KOHDEN CORPORATION
attn: shama mooman
1-31-4 nishiochia
shinjuku-ku, tokyo 161-8 560
JA  161-8560
Manufacturer (Section G)
NIHON KOHDEN TOMIOKA CORPORATION
attn: shama mooman
1-1 tajino
tomioka city, gunma 370-2 314
JA   370-2314
Manufacturer Contact
shama mooman
seibu bldg 2, 4th floor 1-11-2
seibu bldg 2, 4th floor 1-11-2
kusunokidai tokorozawa, saitama 359-8-580
JA   359-8580
MDR Report Key13462856
MDR Text Key288710277
Report Number8030229-2022-02434
Device Sequence Number1
Product Code MHX
UDI-Device Identifier04931921114131
UDI-Public04931921114131
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K102376
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type User Facility
Reporter Occupation Biomedical Engineer
Type of Report Initial,Followup
Report Date 04/19/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received02/04/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberCNS-6201A
Device Catalogue NumberPU-621RA
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received06/17/2022
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured12/26/2013
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-