Additional information was added to d9, h3, h4, h6 and h10.H4: the lot was manufactured from april 23, 2021 - april 24, 2021.H10: two (2) actual samples were received for evaluation.Visual inspection was performed using the naked eye which observed that the housing was malformed which caused the coil cap to separate from the housing.This would render the device unusable and the device would be discarded; the customer has the option to replace with new supplies.The reported condition was verified.The cause of the condition was due to extreme heat temperature during shipping.A batch review was conducted and there were no deviations found related to this reported condition during the manufacture of this lot.Should additional relevant information become available, a supplemental report will be submitted.
|